Best budget · Light meters
Dr.meter LX1330B
by Dr.meter
The classic budget lux meter: 0.1 to 200,000 lux, a separate sensor probe, and independent test data that shows where it errs. About $45 on Amazon.
Specs
| Range | 0.1–200,000 lux |
|---|---|
| Accuracy | ±(4% + 10 dgt) below 20,000 lux |
| Cosine corrected | Not stated |
| Melanopic reading | No |
| Sensor | Separate probe |
| Accuracy high | ±(5% + 10 dgt) at 20,000 lux and above |
| Calibration | 2,856 K incandescent lamp |
| Ranges | 4 manual: 200 / 2,000 / 20,000 / 200,000 lux |
| Display | 3½ digit, max 1999 |
| Sample rate | 2 per second |
| Battery | 9V, about 200 hours |
| Warranty | 1 year |
Specs from the manufacturer and the sources listed below. Last verified Oct 20, 2026.
Good
- Independent tests exist: an unbranded LX-1330B showed a 3.2% calibration offset and 8.3% spread across light sources against a Class A reference
- Separate sensor probe makes it easier to keep your hand out of the light path
- 200,000 lux range covers SAD lamps and most daylight
- Peak hold and data hold buttons
Not so good
- Calibrated to incandescent light; the same test found one 6,500 K LED read 20% high
- Manual range switching, and on the 20,000 lux range you multiply the display by 10
- No cosine correction stated
- Sold out on Dr.meter's own site; Amazon is the practical place to buy
- 1-year warranty, shorter than most meters here
The LX1330B has been sold under many labels for years, and that is its advantage: people have measured it. In a 2018 Budget Light Forum comparison against a MobiLux Class A meter, an unbranded LX-1330B read a 4,000 K LED almost exactly, read another LED 8% low and read a 6,500 K LED 20% high.
The tester later summarized it as a 3.2% average calibration offset with 8.3% spread across sources. That is good for the money, but it means you should treat readings under LEDs as approximate.
The reason is in the manual. Dr.meter calibrates the meter to a 2,856 K incandescent lamp. Under incandescent or halogen light it should be close to spec. Under LEDs, whose spectra differ, the error depends on the LED.
For a SAD lamp check, that error matters less than distance. In a study of light therapy devices, one box measured about 12,000 lux at 20 inches and about 7,000 at 28 inches (Oldham 2019). Measure at eye height with the probe facing the lamp.
Two quirks to know. Ranges are manual, so you must press the range button to reach 20,000 or 200,000 lux. And the display only shows up to 1999, so on the 20,000 lux range a reading of 1000 means 10,000 lux.
Dr.meter’s product page lists accuracy as ±3% ±10 digits, while its PDF manual says ±(4% + 10 digits) below 20,000 lux. The table above uses the more conservative manual figure.
If you want app logging and do not mind a built-in sensor, the UNI-T UT383BT costs about $10 less. If you want cosine correction and a firmer spec, the Extech LT300 is the step up.
Sources
Alternatives
- Best overallExtech LT300$244
- Best for melanopic lightUPRtek MK350S Premium$4,125
- Best for LED lampsExtech LT45$359
- Best with app loggingUNI-T UT383BT$35